User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: HP 5334A/B COUNTER VERIFICATION (OPT 060,030) REV 2.0 DATE: 20-Oct-93 AUTHOR: User Contributed REVISION: 2.0 ADJUSTMENT THRESHOLD: 80% NUMBER OF TESTS: 34 NUMBER OF LINES: 302 CONFIGURATION: Fluke 6060B CONFIGURATION: HP 3325A CONFIGURATION: HP 5335A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is derived from the HP 5334B manual,P/N 05334-90030 # Copyright 1988. The timebases will be locked together so that the frequency # accuracy TUR of the generators has no effect. 1.001 ASK- R N P F L W 1.002 HEAD Preliminary instructions 1.003 DISP Set the HP 5334B TIMEBASE INT/EXT switch to EXT. 1.003 DISP Connect a BNC tee to the HP 3325A EXT REF IN connector. 1.003 DISP 1.003 DISP Connect one side of the tee to the HP 5334B EXT REF IN. 1.003 DISP 1.003 DISP Connect the other side of the tee to the Fluke 6062A 1.003 DISP 10 MHZ OUT connector. 1.003 DISP Connect the HP 5334B to port 1 of the controller. 1.004 HEAD {Frequency Response and Sensitivity Tests, Channel A} 1.005 IEEE [TERM CR] 1.006 IEEE IN 1.007 IEEE [@3325] FR10HZ FU1 AM15MR OF0VO 1.008 DISP Connect the 50 OHM FEEDTHRU TERMINATION to the HP 5334B 1.008 DISP Channel A front panel INPUT connector. 1.008 DISP Connect the HP 3325A SIGNAL output to the FEEDTHRU 1.008 DISP TERMINATION. 1.008 DISP 1.008 DISP Set the HP 5334B Channel A TRIGGER LEVEL/SENS knob 1.008 DISP to the maximum clockwise position. 1.009 IEEE FN1 SE1 RE 1.010 IEEE [D2000] [I] 1.011 MEME 1.012 MEME 10 1.013 MEMC HZ 0.03U 2.001 IEEE [@3325] FR20MH 2.002 IEEE [D1000][I] 2.003 MEME 2.004 MEME 20000000 2.005 MEMC HZ 0.3U 3.001 IEEE AA1 RE 3.002 IEEE [@3325] FR30HZ 3.003 IEEE [D2000][I] 3.004 MEME 3.005 MEME 30 3.006 MEMC HZ 0.03U 4.001 IEEE [@3325]FR20MH 4.002 IEEE [D1000][I] 4.003 MEME 4.004 MEME 20000000 4.005 MEMC HZ 0.3U 5.001 IEEE AZ1 RE 5.002 DISP Remove the 50 OHM FEEDTHRU TERMINATION. 5.002 DISP 5.002 DISP Connect the HP 3325A SIGNAL output directly to the 5.002 DISP HP 5334B Channel A INPUT. 5.003 IEEE [@3325]FR1MH 5.004 IEEE [D1000][I] 5.005 MEME 5.006 MEME 1000000 5.007 MEMC HZ 0.04U 6.001 IEEE [@3325]FR20MH 6.002 IEEE [D1000][I] 6.003 MEME 6.004 MEME 20000000 6.005 MEMC HZ 0.3U 7.001 HEAD {Frequency Response and Sensitivity Tests, Channel B} 7.002 DISP Connect the FEEDTHRU TERMINATION to the HP 5334B 7.002 DISP Channel B front panel INPUT. 7.002 DISP Connect the HP 3325A SIGNAL output to the FEEDTHRU 7.002 DISP TERMINATION. 7.002 DISP 7.002 DISP Set the HP 5334B Channel B TRIGGER LEVEL/SENS knob 7.002 DISP to the Maximum clockwise position. 7.003 IEEE FN2 SE1 RE 7.004 IEEE [@3325] FR10HZ 7.005 IEEE [D1000][I] 7.006 MEME 7.007 MEME 10 7.008 MEMC HZ 0.03U 8.001 IEEE [@3325]FR20MH 8.002 IEEE [D1000][I] 8.003 MEME 8.004 MEME 20000000 8.005 MEMC HZ 0.3U 9.001 IEEE BA1 RE 9.002 IEEE [@3325]FR30HZ 9.003 IEEE [D2000][I] 9.004 MEME 9.005 MEME 30 9.006 MEMC HZ 0.03U 10.001 DISP Remove the FEEDTHRU TERMINATION. 10.001 DISP 10.001 DISP Connect the HP 3325A SIGNAL output directly to the 10.001 DISP HP 5334B Channel B INPUT. 10.002 IEEE BZ1 RE 10.003 IEEE [@3325] FR1MH 10.004 IEEE [D2000][I] 10.005 MEME 10.006 MEME 1000000 10.007 MEMC HZ 0.04U 11.001 IEEE [@3325]FR20MH 11.002 IEEE [D1000][I] 11.003 MEME 11.004 MEME 20000000 11.005 MEMC HZ 0.3U 12.001 IEEE IN 12.002 HEAD {Channel A High Frequency Sensitivity and Response} 12.003 DISP Connect the FEEDTHRU TERMINATION to the HP 5334B 12.003 DISP REAR PANEL Channel A INPUT BNC connector. 12.003 DISP 12.003 DISP Connect the Fluke 6062A RF OUTPUT to the HP 5334B 12.003 DISP FRONT PANEL Channel A INPUT. 12.004 IEEE IN 12.005 IEEE [D2000]FN1 SE1 RE 12.006 6060 8.0e7H -13D -D S 12.007 IEEE [D1000] 12.008 IEEE [D1000][I] 12.009 MEME 12.010 MEMC HZ 2U 13.001 6060 1.00e8H -13D -D S 13.002 IEEE [D1000][I] 13.003 MEME 13.004 MEMC HZ 2U 14.001 IEEE AA1 RE 14.002 IEEE [D1000] 14.003 6060 8.0e7H -13D -D S 14.004 IEEE [D1000][I] 14.005 MEME 14.006 MEMC HZ 2U 15.001 6060 1.00e8H -13D -D S 15.002 IEEE [D1000][I] 15.003 MEME 15.004 MEMC HZ 2U 16.001 HEAD {Channel B High Frequency Sensitivity and Response} 16.002 DISP Remove the FEEDTHRU TERMINATION from the HP 5334B 16.002 DISP REAR PANEL channel A INPUT. 16.002 DISP 16.002 DISP Connect the FEEDTHRU TERMINATION to the HP 5334B 16.002 DISP REAR PANEL Channel B INPUT connector. 16.002 DISP 16.002 DISP Connect the Fluke 6062A RF OUTPUT to the FRONT PANEL 16.002 DISP Channel B INPUT connector. 16.003 IEEE FN2 SE1 RE 16.004 6060 8.0e7H -13D -D S 16.005 IEEE [D2000][I] 16.006 MEME 16.007 MEMC HZ 2U 17.001 6060 1.00e8H -13D -D S 17.002 IEEE [D2000][I] 17.003 MEME 17.004 MEMC HZ 2U 18.001 IEEE BA1 RE 18.002 6060 8.0e7H -13D -D S 18.003 IEEE [D2000][I] 18.004 MEME 18.005 MEMC HZ 2U 19.001 6060 1.00e8H -13D -D S 19.002 IEEE [D2000][I] 19.003 MEME 19.004 MEMC HZ 2U 20.001 HEAD {Period Test} 20.002 DISP Remove the FEEDTHRU TERMINATION from the HP 5334B 20.002 DISP REAR PANEL Channel B INPUT. 20.002 DISP 20.002 DISP Connect the FEEDTHRU TERMINATION to the HP 5334B 20.002 DISP REAR PANEL Channel A INPUT. 20.002 DISP 20.002 DISP Connect the Fluke 6062A RF OUTPUT to the HP 5334B 20.002 DISP FRONT PANEL Channel A INPUT. 20.003 6060 1.0e-8T -13D -D S 20.004 IEEE IN 20.005 IEEE [D2000]FN4 GA1.0 SE1 RE 20.006 IEEE [D500][I] 20.007 MEME 20.008 MEME 1e-008 20.009 MEMC Seconds 1U 21.001 HEAD {Pulse Width Test} 21.002 6060 1.0e-8T -1D -D S 21.003 IEEE IN 21.004 IEEE [D4000]FN10 RE 21.005 IEEE [D1000][I] 21.006 MEME 21.007 MEME 5e-009 21.008 MEMC Seconds 80% 22.001 HEAD {Time Interval A To B Test} 22.002 DISP Disconnect the Fluke 6062A from the HP 5334B. 22.002 DISP 22.002 DISP Connect the HP 3325A SIGNAL output to the HP 5334B 22.002 DISP front panel Channel A INPUT. 22.002 DISP 22.002 DISP Remove the FEEDTHRU TERMINATION from the REAR PANEL 22.002 DISP of the HP 5334B. 22.003 IEEE IN 22.004 IEEE [@3325]FU2 AM70.0MR FR5MH 22.005 IEEE FN5 CO1 GV1 SE1 RE 22.006 IEEE AZ1 BS1 RE 22.007 IEEE [D500][I] 22.008 MEME 22.009 MEME 1e-007 22.010 MEMC Seconds 6% 23.001 HEAD {Time Interval A To B Delay Test} 23.002 IEEE [@3325]FU2 23.003 IEEE [@3325][D5000]FR100HZ AM70MR 23.004 IEEE IN 23.005 IEEE [D5000]FN6 GA.009 SE1 RE 23.006 IEEE CO1 AZ1 RE 23.007 IEEE BS1 RE 23.008 IEEE GS1 RE 23.009 IEEE [D2000][I] 23.010 MEME 23.011 MEME 0.015 23.012 MEMC Seconds 0.067% 24.001 HEAD {Ratio A/B Test} 24.002 IEEE IN 24.003 DISP Connect the HP 3325A SIGNAL output to the HP 5334B 24.003 DISP Channel B INPUT. 24.003 DISP 24.003 DISP Connect the Fluke 6062A RF OUT to the Channel A INPUT. 24.004 IEEE [@3325]FR5MH AM100MR FU1 24.005 IEEE FN7 SE1 AZ1 BZ1 RE 24.006 6060 1.0e7H -10D -D S 24.007 IEEE [D2000][I] 24.008 MEME 24.009 MEME 2 24.010 MEMC RATIO 1U 25.001 HEAD {Rise/Fall Time A Test} 25.002 DISP Disconnect the Fluke 6062A from the Channel A INPUT. 25.002 DISP 25.002 DISP Disconnect the HP 3325A SIGNAL output from the HP 5334B 25.002 DISP Channel B INPUT, and connect it to the Channel A INPUT. 25.003 IEEE IN 25.004 IEEE [@3325]FR10MH AM177MR FU1 25.005 IEEE FN11 AZ1 RE 25.006 IEEE [D1000][I] 25.007 MEME 25.008 MEME 3e-008 25.009 MEMC Seconds 33% 26.001 IEEE AS1 RE 26.002 IEEE [D500][I] 26.003 MEME 26.004 MEME 3e-008 26.005 MEMC Seconds 33% 27.001 IEEE [@3325]FR100HZ AM707MR FU5 27.002 IEEE [D5000][I] 27.003 MEME 27.004 MEME 0.008 27.005 MEMC Seconds 8% 28.001 IEEE [@3325]FU4 28.002 IEEE AS0 RE 28.003 IEEE [D5000][I] 28.004 MEME 28.005 MEME 0.008 28.006 MEMC Seconds 8% 29.001 HEAD {Input C Tests, option 030} 29.002 IEEE IN 29.003 DISP Disconnect the HP 3325A SIGNAL output from the HP 5334B 29.003 DISP Channel A INPUT. 29.003 DISP 29.003 DISP Connect the Fluke 6062A RF OUTPUT to the HP 5334B 29.003 DISP REAR PANEL INPUT C connector. 29.004 6060 9.0e7H -23.5D -D S 29.005 IEEE FN3 RE 29.006 IEEE [D500][I] 29.007 MEME 29.008 MEMC HZ 2U 30.001 6060 2.00e8H -23.5D -D S 30.002 IEEE [D500][I] 30.003 MEME 30.004 MEMC HZ 3U 31.001 6060 1.000e9H -9.5D -D S 31.002 IEEE [D500][I] 31.003 MEME 31.004 MEMC HZ 20U 32.001 6060 1.300e9H -9.5D -D S 32.002 IEEE [D500][I] 32.003 MEME 32.004 MEMC HZ 20U 33.001 HEAD {Timebase Accuracy Test, Standard Oscillator} 33.002 DISP Remove the connection from the HP 5334A/B TIMEBASE. 33.002 DISP 33.002 DISP Connect the HP 5334A/B TIMEBASE IN/OUT to the INPUT A 33.002 DISP of the HP 5335A COUNTER. Set the HP 5334A/B TIMEBASE 33.002 DISP to INT. (internal) 33.003 IEEE [@5335]IN 33.004 IEEE [@5335]FN1 AZ1 AA1 RE 33.005 IEEE [@5335][D3000][I] 33.006 MEME 33.007 MEME 10000000 33.008 MEMC HZ 30U 34.001 HEAD [27][91]1mAUTOMATED VERIFICATION COMPLETE![27][91]m 34.002 DISP The interval is 245 Days. 34.003 END